Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01483-1
Product Name/Description Balt Cristal Balloons

Product Number: CBV18x40/110

Batch Number: 00294780 - Expiry Date: 30/09/2022

ARTG Number: 164958
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 4/12/2017
Responsible Entity Pyramed Pty Ltd
Reason/Issue Pyramed has identified a potential problem with the Y Connector detaching from the outer sheath of the device.

This may result in the balloon remaining partially inflated in the patient and this may require manual and/or surgical retrieval.
Recall Action Recall
Recall Action Instructions Pyramed is advising customers to isolate affected stock from inventory. Pyramed will arrange for collection and replacement of affected devices.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.