Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01450-1 |
Product Name/Description |
C- Series lifts
Manufactured between April 25, 2017 and June 12, 2017
ARTG 292642 |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
13/12/2017 |
Responsible Entity |
|
Reason/Issue |
During internal cycle testing of C-Series lifts Handicare has found that the C-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop. This issue was identified internally. To date, CHS is not aware of any C-Series lift strap incidence related to this issue. |
Recall Action |
Recall |
Recall Action Instructions |
CHS Healthcare will be in contact with affected customers to arrange removal of the current hoist and reinstall of a new unit.
This action has been closed out on 26/02/2019 |
Contact Information |
1300 789 420 - CHS Healthcare |