Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01450-1
Product Name/Description C- Series lifts

Manufactured between April 25, 2017 and June 12, 2017

ARTG 292642
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 13/12/2017
Responsible Entity CHS Healthcare Pty Ltd
Reason/Issue During internal cycle testing of C-Series lifts Handicare has found that the C-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
This issue was identified internally. To date, CHS is not aware of any C-Series lift strap incidence related to this issue.
Recall Action Recall
Recall Action Instructions CHS Healthcare will be in contact with affected customers to arrange removal of the current hoist and reinstall of a new unit.

This action has been closed out on 26/02/2019
Contact Information 1300 789 420 - CHS Healthcare