Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01441-1
Product Name/Description Thermo Scientific Remel Shigella sonnei (Phases 1& 2) Agglutinating Serum. An in vitro diagnostic medical device (IVD)

Product code: R30164201

Lot Numbers: 1465529, 2118879, 2173683, 2161363, 2149414, 2135578, 1738564, 1957305, 1947930, 1847285

ARTG Number: 236824
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/11/2017
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue Thermo Fisher Scientific, has confirmed that Shigella sonnei (Phases 1& 2) Agglutinating Serum may fail to agglutinate within the specified minimum reaction time when tested with Shigella sonnei phase 2 bacteria. Continued use of these lots may result in a failure to correctly identify isolates as Shigella sonnei phase 2.
The products continue to correctly identify Shigella sonnei phase 1 bacteria.
Recall Action Recall
Recall Action Instructions Thermo fisher is advising users to inspect stock and remove any remaining units of the affected batches from use. A review of previously generated patient results is at the discretion of the Laboratory Manager.

This action has been closed out on 19/11/2018
Contact Information 08 8238 9044 - Thermo Fisher Scientific