Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01423-1
Product Name/Description Spectra System/Legacy Guided Surgery Handle Kit Inserts

Part Numbers: G-HK

Kit Lot Numbers: 42562, 52987, 64245 and 74772

ARTG Number: 296286
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 20/11/2017
Responsible Entity Implant Direct Oceania Pty Ltd
Reason/Issue Implant Direct Oceania Pty Ltd is informing users of a recall of four (4) lots of Spectra System/Legacy Guided Surgery Handle Kit/Inserts due to receiving 3 complaints on these products. Through the reporting process, Implant Direct have found that the G-HK Guided Surgery Handle inserts may be out of Implant Direct specification. The 3.4mm guide insert will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide may be too large for the drill. These discrepancies may lead to the possibility of a delay in procedure for the patient if the drilling sequence is discontinued or the clinician would need to continue the surgical procedure without the use of the drill guide.
There have been no injuries as a result of this issue.
Recall Action Recall
Recall Action Instructions Implant Direct is advising users to inspect stock and return affected units.

This action has been closed out on 24/03/2020