Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01414-1
Product Name/Description Foetal Haemoglobin Monoclonal Antibody (HBF-1). An in vitro diagnostic medical device (IVD)

Catalogue Number: MHFH01 and MHFH04

Lot Number: 1848553A and 1781432A

Expiry date: 31 January 2019

ARTG Number: 212744
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 15/11/2017
Responsible Entity Life Technologies Australia Pty Ltd
Reason/Issue Life Technologies Australia have received four (4) customer complaints worldwide on multiple lots pertaining to poor antibody detection whereby customers reported little to no antibody staining. In the complaints received, customers identified a reduction in the intensity of staining, unequal staining, and in some instances, may have observed a reduction or shift in the positive population peak for foetal haemoglobin. With the use of controls, the end user can identify a product deficiency before diagnostic samples are analysed. Failure of the positive controls to perform as expected is a basis for discarding test results and repeating the test.
There have been no reports of injuries as a result of this issue.
Recall Action Recall
Recall Action Instructions Life Technologies is advising users to inspect stock and discard any remaining units of the affected lot number. There is unaffected stock available.

This action has been closed out on 3/09/2018.
Contact Information 1800 636 327 - Life Technologies Australia Pty Ltd