Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01411-1
Product Name/Description i-STAT 1 Analyser Software with i-STAT Total â-hCG Cartridge with software version JAMS 143/CLEW A34. An in vitro diagnostic medical device (IVD)

ARTG Number: 203349
(Abbott Australasia Pty Ltd (Point-of-Care Division) - Clinical chemistry hormone IVDs)
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 13/11/2017
Responsible Entity Abbott Australasia Pty Ltd Point-of-Care Division
Reason/Issue Through a combination of internal studies and customer reports, Abbott Point of Care has determined that users may experience an increased frequency of Quality Check Code 149 when testing i-STAT Total ß-hCG cartridges with i-STAT 1 Analyser software version JAMS 143 / CLEW A34.

There is no impact on user/patient safety. Due to the fact that Quality Check Code 149 requires the test to be repeated, this issue has the potential to delay generation of i-STAT Total ß-hCG results.
Recall Action Recall for Product Correction
Recall Action Instructions Abbott is advising facilities performing testing with i-STAT Total ß-hCG cartridges, to verify that all i-STAT analysers performing i-STAT Total ß-hCG testing have been updated to software version JAMS 144 / CLEW A35. If the software has not yet been updated to version JAMS 144 / CLEW A35, please follow the directions on the Abbott Point of Care website (see link below) for the installation of the software.

https://www.pointofcare.abbott/us/en/offerings/support/software-license-keys/istat-system-software-update .

This action has been closed out on 31/05/2019
Contact Information 1800 816 696 - Abbott Point of Care Technical Support