TGA Recall Reference RC-2017-RN-01405-1
Product Name/Description Bard ArcticSun ArticGel Pads

Item Code: 31700
Lot/Serial Numbers: NGAW3898 and NGBP1302

Item Code: 31705
Lot/Serial Numbers: NGAZ2312 and NGBN2492

Item Code: 31707
Lot/Serial Numbers: NGBN3536

ARTG Number: 262560 (Heating unit, pad, water)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/11/2017
Responsible Entity Bard Australia Pty Ltd
Reason/Issue The manufacturer, Bard Medical Division (BMD), has identified specific lots of Bard ArcticSun ArticGel Pads that have insufficient adhesion of the hydrogel to the pad, which has the risk of the hydrogel peeling away from pads during the removal of the release liner.
Recall Action Recall for Product Correction
Recall Action Instructions Bard is providing customers with additional instructions for use to identify the potential issue, and provide guidance to allow the product to be used in a normal manner.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.