Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01393-1
Product Name/Description Haemoglobin A1c (HbA1c) Flex reagent cartridge used with Dimension Vista with software version 3.6.2 or higher. An in vitro diagnostic medical device (IVD)

Assay: Dimension Vista HbA1C

Catalogue Number: K3105A

Siemens Material Number: 10470481

ARTG Number: 181689
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/11/2017
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare Diagnostics has previously confirmed that when whole blood samples are run from the Small Sample Container (SSC), the HbA1c value may contain a percent bias range of - 8.2% to + 6.5% (TGA Ref: RC-2014-RN-00922-1). Due to this bias, use of the SSC was withdrawn. As a follow up Siemens Healthcare Diagnostics has confirmed that the SSC can now be used with the Dimension Vista Haemoglobin A1c (HbA1c) assay with software version 3.6.2 or higher. The HbA1c Instructions For Use beginning with lot 17205BA will detail use of the SSC.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users that samples for the HbA1c method can be assayed from a Small Sample Container (SSC) loaded in a “Surplus” sample rack and processed as Routine or STAT.
When using an SSC, pipette a minimum of 400 uL of the whole blood sample and load in a “Surplus” sample rack using a 13mm tube insert (Do not run from a “Limited” sample rack).

This action has been closed out on 3/09/2018.
Contact Information 1800 310 300 - Siemens Technical Support Centre