TGA Recall Reference RC-2017-RN-01371-1
Product Name/Description Drystar DT 2 C Hardcopy Film 14" x 17"

Reference: EKMCU

Batch Numbers: 36650011 (exp 2018-06) and 69030031 (exp 2018-10)

ARTG Number: 119352 (AGFA HealthCare Australia - Photographic film, diagnostic imaging)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/10/2017
Responsible Entity AGFA HealthCare Australia
Reason/Issue Agfa Healthcare is advising that due to an error in manufacturing, specific film boxes have unreadable RF-tags. If these boxes are used in printers, the printer will generate an error message, and it will not be possible to print images.

If users were able to use the affected film there is no problem with the image quality or the diagnostic properties.
Recall Action Recall for Product Correction
Recall Action Instructions Agfa Healthcare is advising users to use an alternative film if the issue occurs.

This action has been closed out on 17/10/2018