Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01353-1
Product Name/Description WECK Auto Endo5 Automatic Hem-o-lok Clip Appliers

Product Code: AE05ML

Lot Numbers: 73F1600689, 73F1600694, 73G1600431, 73H1600666, 73H1600667, 73H1600668

ARTG Number: 126369 (Teleflex Medical Australia Pty Ltd - Applier, surgical, clip, haemostatic, single use)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/10/2017
Responsible Entity Teleflex Medical Australia Pty Ltd
Reason/Issue Teleflex has identified certain lots of WECK Auto Endo5 Automatic Hem-o-lok Clip Appliers have a nonconformity that may cause clips to misload, jam, or fall out of the applier. There could be a delay in procedure if an applier becomes jammed or if a clip falls out of an applier.
Recall Action Recall
Recall Action Instructions Teleflex is advising users to inspect stock and quarantine all units from listed product and lot numbers. Teleflex will arrange collection of product from affected users.

This action has been closed out on 07/08/2019