Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01346-1
Product Name/Description SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK

Product Code/Part Number: D684-00-0576-01

Lot Number: 3000043375

ARTG Number: 139645
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/11/2017
Responsible Entity Getinge Australia Pty Ltd
Reason/Issue During internal pressure (leak) testing of the affected product, failures related to the tip seal leak were identified by Getinge. This issue at the catheter tip seal could potentially cause a leak in the balloon and may lead to an interruption of therapy resulting in the need to replace the defective IABC with a new catheter. Such occurrences could cause momentary haemodynamic instability in a patient, or potential complications related to vessel bleeding related to replacement of the IAB catheter.

To date, there have been no reports of serious injury.
Recall Action Recall
Recall Action Instructions Getinge is advising users to quarantine any remaining stocks of the affected product for return.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.
Contact Information 1800 438 464 - Getinge Customer Service