Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01314-1
Product Name/Description Ingenuity TF PET/CT System with software version 4.0.3.26129

ARTG Number: 292543
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 17/10/2017
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue Philips is advising of an issue with the QC Baseline Test. During the PET Daily Quality Control (QC) Baseline Test, customers may see a popup message on the console stating “Communications Error. Do you wish to continue?”, with the options of “OK” or “Cancel”. Answering either way may cause the Daily QC to fail. In some cases, once this message appears, it will re-appear each time the Daily QC is run.
Recall Action Recall for Product Correction
Recall Action Instructions Where Daily QC is unable to be run due to a continual Communication Error, Philips is advising customers to contact their Philips Service Representative to resolve the issue. Philips will correct the issue in an upcoming software release.

This action has been closed out on 18/11/2020
Contact Information 1800 251 400 - Philips Customer Care Centre