Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01286-1 |
Product Name/Description |
Ceiling Unit CU5001/3 and CU 3-21/23
Product Code: CU 5001 Reference Number: 604 105 Serial Numbers: from P51 up to 274
Product Code: CU 5003 Reference Number: 604 106 Serial Numbers: from P51 up to 127
Product Code: CU 3-21 Reference Number: 604 650 Serial Numbers: from P51 up to P58
Product Code: CU 3-23 Reference Number: 604 700 Serial Numbers: from P51 up to P60
ARTG Number: 267878 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
13/10/2017 |
Responsible Entity |
|
Reason/Issue |
The microscope arm of specific ceiling units can be electrically driven from the parked position into the operating field by an electromotive lifting column. The manufacturer Moller-Wedel has identified that abrasive wear of the drive assembly can occur which could cause a slide-through of the telescopic column into the lowest stop.
The microscope arm can be moved upwards and downwards with a levitation height of 40cm by a motorised adjusted telescopic column. If severe abrasive wear occurs the lifting column can slide downwards during lifting procedures up to 40cm onto the lowest stop. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Device Technologies is advising users to move the telescope column to the lowest position for use without restrictions. Users should attach provided labels to the ceiling unit to indicate that the lifting column is in the lowest position until service has been performed.
This action has been reviewed and is considered complete in accordance with the requirements of the URPTG. |