Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01286-1
Product Name/Description Ceiling Unit CU5001/3 and CU 3-21/23

Product Code: CU 5001
Reference Number: 604 105
Serial Numbers: from P51 up to 274

Product Code: CU 5003
Reference Number: 604 106
Serial Numbers: from P51 up to 127

Product Code: CU 3-21
Reference Number: 604 650
Serial Numbers: from P51 up to P58

Product Code: CU 3-23
Reference Number: 604 700
Serial Numbers: from P51 up to P60

ARTG Number: 267878
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/10/2017
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue The microscope arm of specific ceiling units can be electrically driven from the parked position into the operating field by an electromotive lifting column. The manufacturer Moller-Wedel has identified that abrasive wear of the drive assembly can occur which could cause a slide-through of the telescopic column into the lowest stop.

The microscope arm can be moved upwards and downwards with a levitation height of 40cm by a motorised adjusted telescopic column. If severe abrasive wear occurs the lifting column can slide downwards during lifting procedures up to 40cm onto the lowest stop.
Recall Action Recall for Product Correction
Recall Action Instructions Device Technologies is advising users to move the telescope column to the lowest position for use without restrictions. Users should attach provided labels to the ceiling unit to indicate that the lifting column is in the lowest position until service has been performed.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.
Contact Information 03 9371 0973 - Device Technologies Australia