Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01269-1
Product Name/Description Infinity Dual Hemo MCable Pod used with the Infinity Acute Care System (IACS) Monitoring Solution

Dual Hemo MCable Pods with Revision Index RI15 and RI16

Manufactured between August 2015 to August 2016

ARTG Number: 138056
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 4/10/2017
Responsible Entity Draeger Australia Pty Ltd
Reason/Issue Dräger has determined through internal testing that Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dräger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.

Dräger has not received any complaints and is not aware of any negative patient consequences arising from this issue. Revisions prior to RI 15 and after RI17 continue to meet the IPX4 standard.
Recall Action Recall for Product Correction
Recall Action Instructions Dräger are advising users to inspect systems to identify any affected products. Dräger will be providing customers with a sleeve to apply in order to prevent the issue from occurring.

This action has been closed out on the 09/10/2018.
Contact Information 03 9244 7248 - Draeger Medical Australia