TGA Recall Reference RC-2017-RN-01261-1
Product Name/Description SOMATOM go.Up and SOMATOM go. Now

Catalogue Numbers: 11061628 and 11061618

ARTG Number: 142119
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 3/10/2017
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare have identified that the bolts on the patient table have been tightened with an inadequate torque, which may lead to a loose nut on the bolt. Under unfavourable conditions several bolts may gradually loosen at the same time having an effect on the stability of the table.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users that if there is any unusual noise or vibration to cease use of the CT system and contact Siemens immediately. Siemens will be correcting all affected units.

This action has been reviewed and is considered complete in accordance with the requirements of the URPTG.