Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01261-1 |
Product Name/Description |
SOMATOM go.Up and SOMATOM go. Now
Catalogue Numbers: 11061628 and 11061618
ARTG Number: 142119 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
3/10/2017 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare have identified that the bolts on the patient table have been tightened with an inadequate torque, which may lead to a loose nut on the bolt. Under unfavourable conditions several bolts may gradually loosen at the same time having an effect on the stability of the table. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is advising users that if there is any unusual noise or vibration to cease use of the CT system and contact Siemens immediately. Siemens will be correcting all affected units.
This action has been reviewed and is considered complete in accordance with the requirements of the URPTG. |
Contact Information |
1800 310 300 - Siemens Customer Care Centre |