Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01233-1
Product Name/Description Oxford Dipper Hoist

Models: OP10013, OP10014, X10013 and X10014

Purchased between 2001 and 2010

Multiple Serial Numbers affected

ARTG Number: 101030
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 18/10/2017
Responsible Entity Sunrise Medical Pty Ltd
Reason/Issue The manufacturer Joerns Healthcare has revised the service life statement and related manufacture guidance for the Oxford Dipper Hoist. It is possible that issues such as corrosion can occur rapidly in units older than 7 years and this may not be externally visible to users. Joerns Healthcare strongly recommends load testing for the Oxford Dipper is carried out at each and every 6 monthly inspection due to the hostile environment in which it resides and operates. The humid and corrosive nature of the pool-side locations can potentially lead to product deficiencies that may otherwise remain undetected.
Recall Action Recall for Product Correction
Recall Action Instructions Load testing for the Oxford Dipper should be carried out at every six-monthly inspection
The Oxford Dippers that are older than 7 years need to be changed out prior to the following scheduled (6 monthly) inspection to ensure the ongoing safety of users.
Recipients are requested to contact all customer(s) supplied with the supplied serial numbers, then complete the supplied Receipt Form with the appropriate box checked, according to disposition of each hoist as soon as the information becomes available or the change out is complete.

This action has been closed out on 06/02/2019