Type of Product Medical Device
TGA Recall Reference RC-2017-RN-01216-1
Product Name/Description Progressa Bed System

Model Number: P7500A

Production/Distribution Dates: October 2013 to April 2017

ARTG Numbers: 216681, 221346, 221363
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/09/2017
Responsible Entity Hill-Rom Pty Ltd
Reason/Issue Hill-Rom Australia (Welch Allyn) has become aware of an issue with the Progressa bed in regions where high mains input supply power may cause the fuses on the power supply PC board to open. The fuses on the power supply PC board opening may cause loss of bed functions.

The Emergency CPR release can still be activated manually to lower the head section to horizontal.
Recall Action Recall for Product Correction
Recall Action Instructions Hill-Rom will be correcting all affected products.

This action has been closed out on 05/11/2018
Contact Information 1800 445 576 - Hill-Rom Technical Support