Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01216-1 |
Product Name/Description |
Progressa Bed System
Model Number: P7500A
Production/Distribution Dates: October 2013 to April 2017
ARTG Numbers: 216681, 221346, 221363 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
27/09/2017 |
Responsible Entity |
|
Reason/Issue |
Hill-Rom Australia (Welch Allyn) has become aware of an issue with the Progressa bed in regions where high mains input supply power may cause the fuses on the power supply PC board to open. The fuses on the power supply PC board opening may cause loss of bed functions.
The Emergency CPR release can still be activated manually to lower the head section to horizontal. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Hill-Rom will be correcting all affected products.
This action has been closed out on 05/11/2018 |
Contact Information |
1800 445 576 - Hill-Rom Technical Support |