Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-01183-1 |
Product Name/Description |
DiaMed Q.C. System. An in vitro diagnostic medical device (IVD)
Material Number: 45950
Lot Number IHD: 45950 46 1 Lot Number SAP: 166035461 Expiry: 25/09/2017
Lot Number IHD: 45950 47 1 Lot Number SAP: 172383471 Expiry: 09/10/2017
ARTG Number: 232955 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
13/09/2017 |
Responsible Entity |
|
Reason/Issue |
Bio-Rad Laboratories have been made aware that samples of the product DiaMed Q.C. System, (IHD lot 45950 46 1 and 45950 47 1) show excessive haemolysis in samples 1 and 4 and weakly in sample 2.
Haemolysed cells may continue to be used as long as they give the expected results. Internal tests performed by Bio-Rad have confirmed that a clear result is returned with manual testing and on the Saxo-Swing, IH-1000 with the haemolysed QC samples. With IH-500 you may observe uninterpretable results on reverse grouping. In this case, please use another sample to perform the reverse grouping. Note: tests involving serum 1 and serum 2 are in no way affected.
The current situation has shown that the haemolysed lots will still allow the customers to perform QC testing for the following applications: - Typing ABO/D and phenotype Rh Kell - Antibody screening and identification - Compatibility testing |
Recall Action |
Recall |
Recall Action Instructions |
1. Distribute this information with in your facility to all those who need to be aware of this notice 2. Bio-Rad kindly requests you to discard all remaining kits of these affected lots, once replacements have been received. Bio-Rad will provide a replacement once the supplied response form is received.. 3. Complete and return the supplied customer response form to Bio-Rad Laboratories as acknowledgment of receipt of this important information, even in the absence of stock. 4. If you have forwarded any product within your facility, please provide a copy of this letter to them and advise Bio-Rad Laboratories.
This action has been closed out on the 25/09/2018. |
Contact Information |
1800 224 354 Bio-Rad Laboratories |