Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00992-1
Product Name/Description Gentian Cystatin C Calibrator Kit. An in vitro diagnostic medical device (IVD)

Part number: A52763

Kit Lot Numbers: 0006 and 0008

ARTG Number: 261302
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/08/2017
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue The manufacturer Gentian has identified an issue with two specific lots of Gentian Cystatin C Calibrator Kit (lot 0006 and 0008). The identified lots of calibrators resulted in a high bias when used on a DxC instrument. Potential risks associated with the increased bias are an underestimation of the Cystatin C value, which again gives an overestimation of GFR. The bias should be detected when using controls for validation of the calibration curve. As a worst case in the event this is not detected by the controls, the bias may be close to 20%.
Recall Action Recall
Recall Action Instructions Beckman Coulter is requesting users:
1. Inspect their stocks and immediately stop using calibrator lots 0006 and 0008;
2. Share this information with laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, provide them a copy of this letter;
3. Complete and return the Response Form supplied with the Customer Letter within 10 days as confirmation of receipt of this important communication; and
4. If you have any questions regarding this notice, contact your local Beckman Coulter Representative.

This action has been closed out on 15/11/2018
Contact Information 1800 060 881 - Beckman Coulter Customer Support Centre