Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00916-1
Product Name/Description Human IgM kit for use on SPAPLUS

Item Number: TBSNK012S

Lot Numbers: 400872 and 407480

ARTG Number: 202167
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/08/2017
Responsible Entity In Vitro Technologies Pty Ltd
Reason/Issue An investigation by In Vitro Technologies Investigations of specific lots of Human IgM kit for use on SPAPLUS have identified that the calibration stability has decreased in comparison to historical released kits. Users will be able to identify when the calibration is no longer valid, as control values will fall outside of the provided acceptable limits.
Recall Action Recall for Product Correction
Recall Action Instructions The Binding Site recommends the following:
1. Batch sample analysis, when possible;
2. Recalibrate the assay when QC results have deviated outside of the specified range;
3. Contact In Vitro Technologies for any further guidance or additional kit requirements; and
4. Complete the acknowledgement form supplied with the customer letter, even if you do not have any of the affected product, and return it by email to QRA@invitro.com.au

This action has been closed out on 25/09/2018
Contact Information 1300 552 003 - In Vitro Diagnostics Customer Care