Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00912-1
Product Name/Description ImmunoCAP Allergen f205, Herring. An in vitro diagnostic medical device (IVD)

Product Number: 14-4837-01

Lot Number: 999G5

ARTG Number: 186054
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 2/08/2017
Responsible Entity Abacus ALS Pty Ltd
Reason/Issue Abacus ALS is advising that ImmunoCAP Allergen f205, Herring, lot 999G5 contains the wrong allergen. This could cause erroneous test results, which may cause a delay in a proper diagnosis, however, the probability of a serious adverse health consequence or serious deterioration in state of health due to a delayed diagnosis is estimated to be negligible.
Recall Action Recall
Recall Action Instructions Actions to be taken by the customer/user:
1. Stop using the affected lots;
2. Discard any unused product;
3. Retest any samples that have been tested using the affected product; and
4. Fill in the Medical Device Recall return response form supplied with the customer letter and return to Abacus ALS.

This action has been closed out on the 16/10/2018.
Contact Information 1800 222 287 - Abacus ALS