Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-00864-1 |
Product Name/Description |
Medtronic Spinous Process Clamps
Spine Referencing Instrumentation
Model/Catalogue Numbers
9734715 Spinous Process Clamp, Tall 9734716 Spinous Process Clamp, Short 9734723 Double Spinous Process Clamp, Tall 9734724 Double Spinous Process Clamp, Short 9734715K Spinous Process Clamp, Tall 9734716K Spinous Process Clamp, Short 9734723K Double Spinous Process Clamp, Tall 9734724K Double Spinous Process Clamp, Short
9735500 Set (contains 9734715 & 9734716)
ARTG Number: 125074 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
18/07/2017 |
Responsible Entity |
|
Reason/Issue |
Medtronic has become aware of instances where Spine Clamps have been damaged when forced open beyond their intended limits during use and will subsequently not open once attached to a patient’s spinous process. When the Spine Clamp is forced open beyond its intended limits, a component (captive washer) may break off. The washer could then be inadvertently left behind in a patient’s body if the breakage occurs during the procedure. If the washer is missing from the device, the spine clamp cannot be re-opened after placement on the spinous process. Medtronic has received six reports where unintended removal of spinous process occurred when attempting to detach the Spine Clamp. The unintended removal of spinous process can lead to damage of adjacent vertebra and cause premature degradation. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Medtronic is providing an interim workaround for users via the Customer Letter. A new version of the clamps with a design mitigation will be distributed as a permanent correction once available.
This action has been closed out on the 11/10/2018. |
Contact Information |
02 9857 9000 - Medtronic |