Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00858-1
Product Name/Description Heron Toilet and Bath Chair

Article Numbers: 880505-01 and 880505-11

Manufactured between 23rd October 2013 to the 16th December 2015

ARTG Number: 229744
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 11/07/2017
Responsible Entity R82 AU Pty Ltd
Reason/Issue R82 Australia has been informed about breakage of the aluminium gas spring head for the tilting mechanism on the Heron. This particular part was changed to a stainless steel version as of January 2016 and replacement kits are to be provided for replacement on units that still have the aluminium version to eliminate the risk of other breakages.
Recall Action Recall for Product Correction
Recall Action Instructions R82 is undertaking a voluntary measure of sending out stainless steel kits to replace the aluminium units to eliminate the risk of breakages and the potential for related injury. Affected customers are asked to follow the instructions in the customer letter to obtain the new kits.

This action has been closed out on 31/10/2018
Contact Information 02 8213 6666 - R82 AU