Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00842-1
Product Name/Description Iris Hi Lo Bed
(adjustable electric bed)

Model/Code Number: PHCBED1

ARTG Number: 197050
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 31/08/2017
Responsible Entity Pride Mobility Products Australia Pty Ltd
Reason/Issue The TGA has determined that the side rails supplied with this bed do not comply with the below listed standard and that there is a resulting risk of patient harm, including but not limited to patient entrapment:
IEC 60601-2-52:2009: Medical electrical equipment - Part 2-52: Particular requirements for the basic safety and essential performance of medical beds
While no safety-related incidents have been reported to date, Pride Australia has, as a precautionary measure have agreed to undertake a Recall for Product Correction.
Recall Action Recall for Product Correction
Recall Action Instructions Pride, requests that users immediately remove the existing side rails in the event they are installed on the bed. To enable you to remove the existing side rails, Pride have supplied a leaflet of instructions demonstrating how to remove the side rails, along with the Customer Letter as an attachment.
Please follow the instructions given to facilitate this recall action.
This action was finalised on 10/09/2018.
Contact Information 03 8770 9600 - Pride Mobility Service Department