Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00819-1
Product Name/Description Trauma Guide Wires, 70cm

Ball Tip Guide Wire 2.4mm
Item Number: 47-2255-008-00

3mm x 70cm Tear Drop Guide Wire
Item Number: 47-2490-098-00

Expiry Date: April 2022

ARTG Number: 118540
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/06/2017
Responsible Entity Zimmer Biomet Pty Ltd
Reason/Issue The design verification for the previous packaging configuration does not cover the 70cm wires. Whilst there has not been any complaints involving a breach in sterility, Zimmer has completed a design verification to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being retrieved from the Market.
Recall Action Recall
Recall Action Instructions Zimmer is requesting that Customers:
1. Review the notification and ensure affected personnel are aware of the contents;
2. Assist your Zimmer Biomet sales representative by quarantining all affected product;
3. A Zimmer Biomet sales representative will remove the affected product from your facility;
4. Complete the Certificate of Acknowledgement as supplied by Zimmer;
4.a Return a digital copy to RAQA.ANZ@zimmerbiomet.com; and
4.b Retain a copy of the Acknowledgement Form with your field action records in the event of a compliance audit of your facilities documentation. This action has been closed-out on 17/08/2018.
Contact Information 02 9483 5246 - Zimmer Biomet