Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00794-1
Product Name/Description BacT/ALERT BPN Culture Bottles. An in vitro diagnostic medical device (IVD)

Reference Number: 279045

Lot Number: 1048256

ARTG Number: 194228
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/06/2017
Responsible Entity Biomerieux Australia Pty Ltd
Reason/Issue BioMérieux has become aware of a crimp defect associated with the BacT/ALERT BPN culture bottles. An anomaly was identified during the manufacturing process that may lead to having excess material at the neck of the bottle. This material may separate from the crimp exposing a sharp edge or causing a minor cut hazard.
Recall Action Recall
Recall Action Instructions Customers are requested to:
- Confirm this letter has been distributed to and reviewed by all appropriate personnel within the organisation.
- Review inventory of BacT/ALERT BPN bottles for Lot Number 1048256 and discard all affected bottles from this lot.
Please store this letter with your bioMerieux instrument documentation.

This action has been closed out on the 16/11/2018.
Contact Information 1800 333 421 - bioMérieux Help Desk Department