Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00736-1
Product Name/Description BVM (Bag-Valve-Mask) Manual Resuscitation Systems

Multiple Reference and Lot Numbers

ARTG Number: 225855
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 20/06/2017
Responsible Entity Intersurgical Australia Pty Ltd
Reason/Issue Intersurgical has identified the possibility that the valve of the BVM Manual Resuscitation System may become stuck during storage. This could result in the
intended tidal volume not being delivered through the valve, which may cause a delay in treatment.
The Instructions For Use (IFU) provided with these products includes the requirement to ensure that the bag, valve and oxygen delivery are functioning correctly prior to use.
Recall Action Recall
Recall Action Instructions Users are asked to inspect their stocks and quarantine all units from the above lot numbers, then complete the attached Reply Form and email or fax it to us so that we may arrange collection and issue credit. Please complete the Reply Form even if you have no stock which is subject to recall, as we require this information to reconcile this process.

This action has been closed out on 31/10/2018
Contact Information 02 8048 3300 - Intersurgical Australia