Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00692-1
Product Name/Description Various Microgenics Application Sheets. An in vitro diagnostic medical device (IVD)

Multiple Assays / Instrument Platforms

ARTG Numbers: 192621 and 235822
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/06/2017
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd
Reason/Issue During an internal assessment, Thermo Fisher Scientific found that Application Sheets provided by the manufacturer Microgenics for expanded use of some assays are inadequate. Specifically, the appropriate validation data for the expanded use of these assays, as indicated in the Instructions for Use (IFU) when not used on Manufacturer approved analysers, is inadequate. In addition, Microgenics did not clearly state in these Application Sheets that the parameters provided are only to be used as starting points and that each assay setting needs to be validated individually by the user.

Thus the information given on the specific Application Sheets for use with non-Microgenics approved analysers may be ineffective. It is possible that Assays not validated for use with other instrument platforms could potentially lead to erroneously low or high test results.
Recall Action Recall for Product Correction
Recall Action Instructions Thermo Fisher Scientific is requesting users to 1) immediately discontinue use of the affected application sheets used on their specific instrument platform, 2) review their internal assay validations to determine if their current instrument settings using these assays require additional action within their clinical laboratory settings (consult a qualified healthcare professional, as appropriate) and 3) contact Customer Service to check if additional action is required.
This recall was finalised on 18/09/2018
Contact Information 03 9757 4300 - Thermo Fisher Scientific