Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00653-1
Product Name/Description GC80 Diagnostic X-Ray system

All Serial/Lot Numbers

ARTG Number: 224370
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/06/2017
Responsible Entity Samsung Electronics Australia Pty Ltd
Reason/Issue An incident had been reported in Germany to Samsung Electronics on 15 Dec 2016 where an image was obtained with over exposure during a thorax examination using the AEC function. The exposure condition was: 125kV, 180mAs.

As a result of the malfunction of the AEC, irradiation may not terminate properly which can cause exposure until the AEC back up time which is set at 500mSec by default. This can cause overexposure for users.
Recall Action Recall for Product Correction
Recall Action Instructions Samsung will be implementing a software update on affected units in early July 2017. In the interim, users are advised that they may continue to use the GC80 system, taking precaution of any possible network disconnection between the THU and WS. Users are requested to check 1) the THU connection status when using the system and 2) the THU connection before logging in after system booting. (Please refer to recall notification for images and further details).

This action has been closed out on 07/11/2018
Contact Information 02 9763 9700 - Samsung Australia