Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00645-1
Product Name/Description Dynabeads HLA Class II and Dynabeads HLA Cell Prep II. An in vitro diagnostic medical device (IVD)

Dynabeads HLA Class II

Catalogue Number: 21003
Lot Numbers: 00461955, 00477414, 00480193

Catalogue Number: 21004D
Lot Numbers: 00461951, 00469604, 00477401, 00480209

Dynabeads HLA Cell Prep II

Catalogue Number: 21903
Lot Number: 00469606, 00477408, 00480204

Catalogue Number: 21904
Lot Numbers: 00455989, 00461965, 00469615, 00492942
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/05/2017
Responsible Entity Bio-Strategy Pty Ltd
Reason/Issue The manufacturer has recently identified an impurity (bacterial endotoxins) in the raw material used for the final formulation of the finished product of Dynabeads HLA Class II and Dynabeads HLA Cell Prep II. Their investigation determined that the presence of these endotoxins along with any deviation from the recommended cell isolation temperature in the user manual (e.g. performing some steps or storing product at room temperature) may reduce cell viability and lead to failure of the negative control.
Recall Action Recall
Recall Action Instructions Bio-Strategy are advising users to discontinue use of the affected material immediately, and to quarantine and place all unused product in a box at ambient temperature ready for collection. Bio-Strategy will arrange for the collection, and will provide replacement product or issue credit..

This action has been closed out on 19/10/2018
Contact Information 1800 008 453 - Bio-Strategy