Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-00627-1 |
Product Name/Description |
Synchron Systems Microalbumin (MA) Reagent
Reference Number: 475100
Lot Number: M609458
Expiry Date: 30 Sep 2018
ARTG Number: 213984 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
15/05/2017 |
Responsible Entity |
|
Reason/Issue |
Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2. As a result, patient results with MA Reagent Lot M609458 may be up to 0.5 mg/dL lower than actual values. In addition, lower recovery of MA in Urine Protein Control Level 1 may cause MA to exceed the lower limits of a laboratory's established QC range. |
Recall Action |
Recall |
Recall Action Instructions |
Beckman Coulter is advising users to discontinue use and discard MA Reagent Lot M609458 following their laboratory's disposal protocol. Replacement product will be provided.
This action has been closed out on the 25/09/2018. |
Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |