Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00627-1
Product Name/Description Synchron Systems Microalbumin (MA) Reagent

Reference Number: 475100

Lot Number: M609458

Expiry Date: 30 Sep 2018

ARTG Number: 213984
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 15/05/2017
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1. Internal testing has confirmed that MA is recovering within the published insert range for Beckman Coulter Urine Protein Control Level 2.
As a result, patient results with MA Reagent Lot M609458 may be up to 0.5 mg/dL lower than actual values. In addition, lower recovery of MA in Urine Protein Control Level 1 may cause MA to exceed the lower limits of a laboratory's established QC range.
Recall Action Recall
Recall Action Instructions Beckman Coulter is advising users to discontinue use and discard MA Reagent Lot M609458 following their laboratory's disposal protocol. Replacement product will be provided.

This action has been closed out on the 25/09/2018.