Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00588-1
Product Name/Description Asserachrom Von Willebrand Factor (VWF):Collagen (CB). An in vitro diagnostic medical device (IVD).

Reference Number: 00239

Lot Number: 114710(RE)

ARTG Number: 185170
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/05/2017
Responsible Entity Diagnostica Stago Pty Ltd
Reason/Issue Following a customer complaint, Stago has investigated and confirmed that Asserachrom VWF:CB lot 114710(RE) gives higher results compared to the other Asserachrom VWF:CB tested lots for samples with a VWF:CB level above 30%. This issue is more pronounced for samples with a VWF:CB level above 50%.

Based on internal testing by Stago, normal samples tested are overestimated by 26% on average (relative deviation). For example: a normal plasma at 134% could be reported at 167%.The pathology samples tested and in particular the type 1 plasmas are overestimated by 38% on average (relative deviation). For example: a type 1 plasma at 56% could be reported at 75%.

The Quality Control (QC) included in the kit does not always detect the issue.
Recall Action Recall
Recall Action Instructions Diagnostica Stago is advising users to stop using and destroy the affected lot according to current local regulations. Replacement kits will be provided.

This action has been closed out on the 25/10/2018.