Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-00578-1 |
Product Name/Description |
Wingman35 Crossing Catheters
Wingman 35 Crossing Catheter, 65cm Reference: WGM35065CE Lot Numbers: 15087074R1
Wingman 35 Crossing Catheter, 90cm Reference: WGM35090CE Lot Numbers:1505144R, 1508264, 1603214R2, 1605194R
Wingman 35 Crossing Catheter, 135cm Reference: WGM35135CE Lot Numbers: 1505264, 1508204, 1604213R, 160620R
ARTG Number: 246377 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
9/05/2017 |
Responsible Entity |
|
Reason/Issue |
The manufacturer ReFlow Medical, has identified specific lots of Wingman35 Crossing Catheters may be prone to tip detachment. If a catheter tip detaches from the Wingman 35 Crossing Catheter within the vasculature, it could cause injuries to blood vessel walls, thrombotic events, foreign body embolisation, heart attacks and death. The manufacturer has received no reports of injury or death associated with the tip failures. |
Recall Action |
Recall |
Recall Action Instructions |
Diverse Devices is advising users to stop using and quarantine all affected products. Affected products are to be returned to Diverse Devices for replacement.
This action has been closed out on 02/11/2018 |