Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00531-1
Product Name/Description Safety Clip - Hoffmann Limb Reconstruction Frame (LRF) Bone Transport

Item Number: 49330401

Lot Numbers: J49236, V07241, V30655, V47575, V49404

ARTG Number: 258035
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/04/2017
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has found that there is the potential for loosening of the safety clips from the intended position on the Quick Release Mechanism of the Transport Struts. In the case of no safety clip and/or overloading, the frame may loosen, bend or break in time, causing frame stability collapse.
Recall Action Recall
Recall Action Instructions Stryker is advising customers to immediately inspect inventory and quarantine affected product. Stryker will coordinate return of the product and provide credit. Clinicians are advised to inspect if the Safety Clips are still in place when patients come in for routine follow up examination and replace the Safety Clips, and to consider contacting patients that do not show up for routine follow up examination. The need to replace the entire Hoffmann LRF Bone Transport if indicated should also be considered.
This action has been closed-out on 21/03/2018.
Contact Information 02 9467 1175 - Stryker