Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00487-1
Product Name/Description SOMATOM Force with software version VA50A

Catalogue Number: 10742326

ARTG Number: 274034
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 19/04/2017
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue An addendum is to be added to the Instructions For Use (IFU) of the SOMATOM Force system with software version VA50A which is related to the use of the Turbo Flash Mode with very short scan ranges. When selecting the Turbo Flash Mode in combination with a spiral scan length of less than 27 mm the image reconstruction will possibly fail for this scan. As a workaround the addendum states for the Turbo Flash Mode to always select a spiral scan length higher than 27 mm or to switch to another mode instead.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users to always select a spiral scan length higher than 27mm or to switch to another mode when selecting the Turbo Flash Mode in combination with a spiral scan length of less than 27mm. (This information is contained in the addendum which will be posted to users). Users are requested to place this addendum in their User Manual and to disseminate the information to the relevant personnel at their organisation.
This action has been closed-out on 23/03/2018.
Contact Information 1800 310 300 - Siemens Customer Care Centre