Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-00481-1 |
Product Name/Description |
ACQUITY UPLC Sample Organisers. An in vitro diagnostic medical device (IVD)
Model Number: 186015020IVD
Lot Numbers: G13UPO544M, J13UPO553M, G13UPO542M
ARTG Number: 201428 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
20/04/2017 |
Responsible Entity |
|
Reason/Issue |
The manufacturer, Waters Corporation has identified an issue with the ACQUITY UPLC Sample Organisers, whereby there is a potential for an internal fuse shield to combust, resulting in a burning odour and instrument failure. An investigation concluded that the potential combustion of the fuse shield would be contained within the instrument with no risk of damage to the samples or the surrounding areas. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Emergo will send a field engineer to attend the user's facility to modify the affected device (by removing the internal fuse). In the interim, users are advised that if they note a burning odour in or around the instrument, to immediately power down the unit, discontinue use, and call their local representative to report the event. This action has been closed-out 16/03/2018. |