Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00455-1
Product Name/Description Visum LED Surgical Lighting system - EDS Light Suspensions, Central Axis

EDS Light Suspensions Item Numbers (Serial Numbers):

0682-001-566 (15K002818, 15K002498, 15K002518, 15K002508)
0682-400-142 (15L004128)
0682-400-148I (16C001798)
0682-001-567I (15L006528)

ARTG Number: 118878
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 19/04/2017
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has become aware that there is a remote likelihood that the EDS Light Suspension, Central Axis may have an insufficiently assembled joint in the suspension. If the EDS Light Suspension, Central Axis falls, it could potentially cause serious injuries to health care providers and/or patients. To date, there have been no adverse events reported to Stryker for this issue worldwide.
Recall Action Recall for Product Correction
Recall Action Instructions A Stryker Technician will contact users to organise a time to perform an inspection of the product. If a correction is required the central axis will be replaced. In the interim, if the product malfunctions, users are to immediately discontinue its use and contact Stryker immediately.
This action has been closed-out on 21/03/2018.