Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00446-1
Product Name/Description BodyGuard Pump IV Administration Set

Catalogue Number: CE-GEN00014

Batch Numbers: 12387350 & 1483529

ARTG Number: 119624
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/04/2017
Responsible Entity REM Systems Pty Ltd
Reason/Issue The manufacturer Caesarea Medical Electronics (CME) has identified that the SR segment in the sets from two batches have slight variations in dimensions, in some places being at borderline specification. These random variations along the tubing, when used in intermittent mode that includes long intervals of zero or near zero rate, may yield under- infusion results.
Recall Action Recall
Recall Action Instructions REM Systems is advising users to review their stock and quarantine the affected batches. Replacement stock will be provided by REM Systems.

This action has been closed out on 13/11/2018