Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00373-1
Product Name/Description cobas 8100 BRF and URF with software versions 02-xx. An in vitro diagnostic medical device (IVD)

cobas 8100 automated workflow series

BRF for cobas 8100
Material Number: 07439954001

URF for cobas 8100
Material Number: 07441657001

ARTG Number: 173887
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/03/2017
Responsible Entity Roche Diagnostics Australia Pty Limited
Reason/Issue An issue regarding the reformatter modules URF and BRF of the cobas 8100 automated workflow series (when used with software version 02-xx) has been identified. This issue is caused by an incorrect mechanical control sequence based on a limitation in software versions 02-xx, resulting in the racks to be shaken and specimen spillage. Sample spillage may cause cross contamination of samples and potentially expose operators to infectious material.
Recall Action Recall for Product Correction
Recall Action Instructions Until the installation of the software upgrade, Roche Diagnostics is advising users to ensure that the rack buffering function is disabled in the configuration of the Control Unit. In the event of a major spillage, users are advised to set the affected modules offline. These modules should be checked and cleaned by a Roche Service Representative before being returned to routine operation.
This action has been closed-out on 15/03/2018.
Contact Information 02 9860 2357 - Roche Diagnostics Australia