Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-00371-1 |
Product Name/Description |
Minop Invent 30 degrees Trocar D 8.3mm L 150mm
Article Code: FH620R
ARTG Number: 106672 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
20/03/2017 |
Responsible Entity |
|
Reason/Issue |
An issue has been identified whereby the insulation of an electrode has been damaged intra-operatively when it is pulled out of the Minop Invent Trocar. Subsequent investigation has shown some burrs at the distal working end of the trocar. |
Recall Action |
Recall |
Recall Action Instructions |
B. Braun is advising customers to complete and return the Reply Form in order for their affected stock to be collected and replaced.
This action has been closed out on the 25/09/2018. |
Contact Information |
1800 251 705 - B. Braun Australia |