Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00365-1
Product Name/Description Cook Bush SL and Bush DL Ureteral Illuminating Catheters
(used for trans-illumination of the ureters during laparoscopic or open surgical procedures)

Cook Bush SL Ureteral Illuminating Catheter Set
Catalogue Numbers: 084100, 084120 , J-BICS-058020
Part Numbers: G16745, G16746, G16737
All Lots

Cook Bush DL Ureteral Illuminating Catheter Set
Catalogue Numbers: 084520, 084510, J-BICS-078000
Part Numbers: G16747, G16143, G16262
All Lots

ARTG Number: 215910
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/03/2017
Responsible Entity William A Cook Australia Pty Ltd
Reason/Issue There has been an increase in reports that the proximal black component that connects to the distal transparent component of the catheter may overheat and melt. There have been no reports of overheating along the transparent component of the catheter, which is the part of the device that comes into contact with the patient’s urethra, bladder and ureter. Potential adverse events that may occur if these devices overheat and come into direct contact with skin include burns to the skin.
Recall Action Recall
Recall Action Instructions Customers are advised to quarantine and immediately return any affected product that remains unused to Cook Medical.

This action has been closed out on 13/11/2018