Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00348-1
Product Name/Description EliA Sample Diluent

Art. No. 83-1023-01

Lot Numbers: JM3Y, JS0B, JP87, JK5N, JU9C

Bottle Lots: BVXDR/JM3W, BVXDT/JS0A, BVXDS/JP83, BVXDP/JK5M, BVXDU/JU9B

ARTG Number: 229543
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/03/2017
Responsible Entity Abacus ALS Pty Ltd
Reason/Issue There have been reports of instrument malfunctions for Phadia 250 due to deformed bottles in the above-mentioned products and lots. The deformation of the bottle can cause an erroneous volume detection by the instrument, which will result in insufficient aspiration of sample diluent, thus generating insufficient dilution of patient samples. The deformation is described as a flange or brim, located on the inside lower part of the vial opening. The instrument malfunction due to deformed bottles could cause erroneous test results, mainly a false positive, with regard to all EliA assays. The frequency of this bottle defect is very low and not every defective bottle will cause the error. An instrument malfunction caused by deformed bottles will generate the instrument error message “3-145 RIGHT ARM LIQUID DETECTION BELOW LOWER LIMIT with Target 2”.
Recall Action Recall
Recall Action Instructions Users are notified to stop using the affected lots and record the lot numbers and quantity of any unused products prior to discarding them. Abacus ALS will organise replacements at no charge. Also, users are advised to discuss the contents of the letter with the Medical Director to ascertain if any further actions are required.
This action has been closed-out on 21/03/2018.
Contact Information 1800 222 287 - Abacus ALS