Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00323-1
Product Name/Description Liaison C-Peptid Assay. An in vitro diagnostic medical device (IVD).

Part Number: 316171

All Lots

ARTG Number: 189435
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/03/2017
Responsible Entity DiaSorin Australia Pty Ltd
Reason/Issue A recent testing of Liaison C-Peptid assay by DiaSorin has highlighted the issue of cross reactivity with human proinsulin. This cross reactivity occurs in 100% of the tested samples and results in an increased C-peptide concentration in all samples tested, across all populations.
For normal patients, the risk of falsely elevated results is rare, since the concentration of proinsulin in healthy fasting people is at least 100 times lower than the C-peptide concentration and therefore has only minimal effect on the results determined with the assay. Patient populations of highest risk for falsely apparently elevated assay results are patients with insulin processing defects, renal malfunction, insulinoma or similar conditions.
Recall Action Recall for Product Correction
Recall Action Instructions DiaSorin is requesting users to confirm receipt of the “Urgent Recall for Product Correction” notification by completing the response section and returning it to DiaSorin. Users are advised of the possibility of further action pending TGA's receipt of additional information.
This action has been closed-out on 23/03/2018.
Contact Information 02 9338 4859 - DiaSorin Australia