Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00318-1
Product Name/Description VisuMax Laser

Serial numbers starting with 963599 up to 1157772

ARTG Number: 150527
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/03/2017
Responsible Entity Carl Zeiss Pty Ltd
Reason/Issue Based on complaints, Carl Zeiss has identified that in certain devices there is an increased probability of failure of the battery test circuit components of the uninterrupted power supply (UPS)-module. It was reported from customers that after treatment was done and machine was switched off white smoke with choking smell came out from the UPS at the rear side of the machine. There were no patient involvement and no adverse consequences associated with these complaints.
Recall Action Recall for Product Correction
Recall Action Instructions The battery circuit for the UPS has been modified to make the UPS more robust. Carl Zeiss will replace all old power supplies as part of the next scheduled service activities of the VisuMax.
This action has been closed-out on 23/03/2018.
Contact Information 02 9020 1333 - Carl Zeiss Pty Ltd