Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00303-1
Product Name/Description Artis Q Systems with A100G generators

Catalogue Numbers: 10848353, 10848354, 10848355, 10848280, 10848281, 10848282, 10848283

ARTG Number: 273952
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 2/03/2017
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue In Artis Systems with A100G generators, a component can be affected by ageing, resulting in the failure of a module in the high-voltage generator. When the failure occurs, the system usually responds by providing emergency fluoroscopy. In the worst case, a spontaneous failure of the radiation generation function is conceivable. In the event of a spontaneous failure, it may be necessary to cancel or restart a clinical treatment or transfer it to a functioning system.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users that potentially affected modules will be replaced. Users are further advised to implement standard emergency processes in the event of a system failure and to have these processes carefully prepared until the affected modules are replaced.
This action has been closed-out on 23/03/2018.
Contact Information 1800 310 300 - Siemens Customer Care Centre