Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00296-1
Product Name/Description LIFEPAK 15 Monitor/Defibrilators with End-Tidal CO2 (EtCO2) functionality

Multiple Serial Numbers

EtCO2 module
Lot Codes: 1506, 1520, 1532, 1544, 1551

ARTG Number: 215648
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 2/03/2017
Responsible Entity Physio-Control Australia Pty Ltd
Reason/Issue An issue has been identified where the EtCO2 reading on the LIFEPAK 15 can intermittently show a value of “XXX” after start-up or during device operation.
An investigation by the manufacturer of the EtCO2 module has revealed that compressed air may have contaminated filters within in the module. As a result of this filter contamination, the EtCO2 display on the device will provide a reading of “XXX”. There may be solid or dashed lines present on the device display in the place of an EtCO2 waveform and the SERVICE LED will be illuminated. In this situation the EtCO2 monitoring functionality will no longer be available.
Recall Action Recall for Product Correction
Recall Action Instructions Physio-Control is advising users who experience the “XXX” after power-up or during device operation to make immediate contact to arrange for servicing of the affected device. Users are further advised that if the issue does not occur, to continue performing the daily device self-tests as indicated in the Operating Instructions and the device is safe to use.

This action has been closed out on 05/11/2018
Contact Information 1800 987 982, Option 2 - Physio-Control