Type of Product |
Medical Device |
TGA Recall Reference |
RC-2017-RN-00233-1 |
Product Name/Description |
Biolox Forte 36mm Alumina Ceramic Head Ceramic Femoral Head 36mm SLT Taper Short Neck Part Number: 26000010
Ceramic Femoral Head 36mm SLT Taper Medium Neck Part Number: 26000011
Ceramic Femoral Head 36mm SLT Taper Long Neck Part Number: 26000012 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
17/02/2017 |
Responsible Entity |
|
Reason/Issue |
An investigation was completed following a complaint by a surgeon that the ceramic femoral head was shorter by 2mm than other 36mm MicroPort femoral heads. MicroPort's investigation confirmed this complaint and concluded that the MicroPort 36mm ceramic femoral heads have been manufactured to this specification for over 15 years. The main complaint issue associated with this finding would be surgical revision for dislocation or laxity of the hip joint. |
Recall Action |
Hazard Alert |
Recall Action Instructions |
Surgical Specialties is advising surgeons to maintain their usual follow-up protocol and actions for their patients and to ensure that these patients are informed about symptoms (particularly pain, instability, difficulty walking and/or performing common tasks) that indicate the need for revision surgery. This action has been closed-out on 22/03/2018. |