TGA Recall Reference RC-2017-RN-00131-1
Product Name/Description NX workstation with NX 3.0.8950 image processing software

ARTG Numbers: 119463, 173861, 178856
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/01/2017
Responsible Entity AGFA HealthCare Australia
Reason/Issue Due to a memory issue, users are shown a warning message about unexpected error and possible incorrect data integrity. A reboot of the workstation computer after having received the warning message prevents the issue from occurring. At this stage, there has been one instance whereby the warning instructions were not followed and resulted in the image of one patient being archived in another patient’s record.
Recall Action Recall for Product Correction
Recall Action Instructions In order to prevent this issue, Agfa HealthCare is advising users to conduct a daily reboot of their workstation by completely turning off the computer and then restarting it. Users are further advised to follow all warning instructions as indicated on their NX workstation and reboot the workstation’s computer if a warning message appears. A new version of the NX software (3.0.9000) will be released shortly and users will receive this upgrade to their installations accordingly.

This action has been closed out on the 12/10/2018.