Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2017-RN-00008-1
Product Name/Description Covidien Devon Light Glove

Product Codes: 571711, 31140208, 31140216, 31140257

All Lot Numbers within expiry beginning with 630XXXXXXX and lower

ARTG Number: 186761
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/01/2017
Responsible Entity Covidien Pty Ltd
Reason/Issue Customers have reported that on rare occasions, the Devon Light Glove may split upon application to the Devon Light Handle Adapter. Some of the reported splits resulted from difficult application of the Light Glove to the Handle Adapter. More recently, clinicians have reported finding splits in the Light Glove following surgery completion, where no difficulty in application of the Light Glove was encountered or finding splits directly out of the package. A split in the Light Glove causes a breach in the sterile field and can increase the potential for infection. There have been two reports in which splits were found at the conclusion of surgery.
Recall Action Recall
Recall Action Instructions Medtronic is advising the distributor to quarantine and discontinue use of the affected lots prior to returning them for a credit. The distributor is further requested to forward the recall letter to hospitals with the affected lots.

This action has been closed out on the 11/12/2018.
Contact Information 02 9857 9000 - Medtronic