Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-01597-1
Product Name/Description BioFlo PICC with ENDEXO and PASV Valve Technology with Nitinol Guidewire - Catheterisation kit, central venous

Description:Bio-Stable 5F DL-55CM IR-145 Kit Valved with Nitinol GW PG

Item Number: H965458830

Catalogue Reference: 45-883

Lot Number: 4908978

Expiry: 30 June 2017

ARTG Number: 203831
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 3/01/2017
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue Navilyst Medical, Inc., the manufacturer of the BioFlo PICC with PASV Valve Technology, is conducting a recall based on information received from Greatbatch Medical, who manufacturers the ViaPeel PTFE Peelable Introducer. Navilyst Medical has confirmed that the affected sheaths from a Greatbatch Medical Recall of the Peelable Introducers (Greatbatch Model Number 10890-006) have been included in packaged BioFlo PICC Kits. These particular Introducers have the potential for the handles to detach from the sheath during use. This may result in user dissatisfaction and/or prolonged procedure, or if both handles detach, the sheath tube could migrate into the patient leading to an unplanned surgical retrieval.. Navilyst Medical has received several complaints from users, but to date, there have been no reports of patient injuries as a result of this issue.
Recall Action Recall
Recall Action Instructions Emergo is advising users to immediately stop using and quarantine the affected lot of Bioflo PICC Kits. Affected product is to be returned to the distributor Florey Medical Pty Ltd. Replacement product or credit will be issued via the distributor.
This action has been closed-out on 19/03/2018.
Contact Information 08 8180 0640 - Florey Medical